產品名稱:Measuring Cardiac Output
產品編號:Measuring Cardiac Output
廠  牌:Transonic Systems, Inc.

PDF下載

Continuous beat-to-beat measurements of mouse cardiac output (minus coronary flow) can be made directly with the new 1.5PSL Nanorobe positioned on the ascending aorta. The flowmeter displays mean flow; recorded waveform data are used in calculations of stroke volume, peak flow, aortic flow acceleration dF/dt, aortic input impedance, systemic vascular resistance and heart rate. Flow measurements with pressure data are used to determine pressure volume relations in cardiac function and can now be applied in transgenic mouse models.

The 1.5PSL flowprobe is designed specifically for chronic implantation on the ascending aorta of the mouse. The transducer housing is < 5 mm and is easily accommodated in the thoracic cavity of the mouse. The nonconstrictive fit does not interfere with the ascending aorta or pulmonary artery. The probe's flexible cable is easily tunneled under the sking for connector exteriorization.
 

    1.5SL_nano
    1.5 PSL

    Chronic Ascending aorta implant protocol in the mouse (ppt)

    1.5 Specifications

    CAT #

    VESSEL

    BIDIRECTIONAL FLOW

    ACCURACY

    ULTRA-SO UND

    Probe

    OD mm
    acute

    Low Flow Normal Flow Max Range
    ml/min ml/min ml/min

    Zero Offset Absolute Relative
    ml/min % %

    Frequency
    MHz

    1.5PSL

    1.0 - 1.5

    10 40 200

    0.2 ± 15 ± 2

    4.8

    conscious_aorta_trace

    Ascending aortic blood flow and pressure in a conscious mouse 7 days after implantation with a 1.5PSL Nanoprobe. Courtesy of B. Janssen, Univ. Maastricht, The Netherlands

     

     

     

    Flowmeter Compatibility

    Transonic System's new 400-Series modular flowmeters are compatible with the 1.5PSL probe. Single channel T106 and dual channel T206 flowmeters manufactured in 2001 that display a mouse sticker ("1.5SL Ready") on the front panel are compatible with the 1.5SL. Older flowmeters cannot be upgraded.
     

    Implantation Techniques

    The same techniques for implantation of Transonic Systems' flowprobes in rats are applied to vessel isolation and probe placement in the mouse, but on a smaller scale. Surgeries on this scale are performed with the aid of a surgical microscope

    Measuring Flow in the Mouse Aorta
    T.L. Smith, PhD, Dept. of Orthopedics, Wake Forest University School of Medicine

    To obtain the detailed procedural video showing step-by-step surgery for implantation of the Transonic 1.5PSL flowprobe on the ascending aorta of the mouse
    To order call: Tel: 607-257-5300; Fax 607-257-7256
    See video stills and implantation protocol below

    Implantation of the 1.5 SL Flowprobe on the Mouse Ascending Aorta

    00000003
    Visualization of ascending aorta in 1.5SL probe lumen.

    00000002
    1.5PSL Flowprobe positioned for closure

    Right Thoracotomy

    Under anesthesia and after the mouse has been properly ventilated on a respirator, a right thoracotomy is performed in the 3rd intercostal space to expose the lungs in the thoracic cavity. The right lung is packed clear of the surgical field with a small piece of surgical sponge. The ascending aorta lies directly under the thymus gland.

    Careful dissection of the ascending aorta is performed by blunt dissection using microsurgical vessel dilators to free the vessel from connective tissue. The vessel can be manipulated most easily by grasping the small fat pad at the base of the pulmonary artery.

    After the aorta has been isolated from the pulmonary artery, two pieces of surgical silk are passed under the vessel to aid in placing the vessel within the lumen of the probe. The flowprobe is introduced into the thoracic cavity and the vessel lifted into the probe lumen so that the J-shaped reflector encircles the aorta.

    After confirmation of vessel placement, the sutures are removed and the 1.5SL flowprobe is rotated so the probe cable exits laterally to the right. The thoracotomy can be closed over the probe to establish negative pressure in the thoracic cavity and improve venous return & cardiac output.

    For acute measurements, an acoustic coupling gel injected into the flowprobe lumen is used to transmit the ultrasound signal. Gel is not required for chronic applications; fibrous tissue encapsulation of the flowprobe will provide good signal transmission after the animal recovers (3-5 days) and is ready for measurement & experimentation.
     

    Summary

    Mouse Cardiac Output & Stroke Volume

    Species: Mouse
    Weight: 20 - 30 grams
    Vessel: Ascending Aorta
    Surgical Approach: Right Thoracotomy
    Duration: Conscious - chronic 3 weeks

    Recommended Probe: 1.5SL
    Cable Length: 4 cm
    Connector: CA4S
    Accessories: Anchor Cuff (AAPC105)

    mouse_aorta
    Ascending aorta waveform of 37.2 gram mouse, 1.5SL probe, 21
    days post-implant. Cardiac output averaged 16.6 ml/min. Heart
    rate was 750 beats/min.

    Cardiac Output: 16.6 ml/min
    Heart Rate: 750 beats/min
    Stroke Volume: 22 microL/min

     

    Significance: Stroke volume used to calibrate impednace catheter for left ventricular relations & studies of cardiac function.

     

    Ascending Aorta 1.5PSL Validation

    Janssen, B., Debets, J., Leenders, P., Smits, J., Chronic Measurement of Cardiac Output in Conscious Mice," AJP, 2002; 282(3) R928-935. 47V

    Gross, V., Luft, FC, "Exercising Retraint in Measuring Blood Pressure in Conscious Mice," Hypertension 2003 41(4) 879-881.

    http://www.transonic.com/cardiac_output.shtml

         
     